CheckMyRecalls

Becton Dickinson & Co. recalls

11 recall notices naming Becton Dickinson & Co. since January 2025, in United States. The most recent was published 21 Oct 2026.

Latest recalls (11)

  1. Becton Dickinson & Co. recalls BD Phoenix Panel NMIC-312 and 1 moreFDAPotential for quality control failures with Klebsiella pneumoniae ATCC BAA-1705 when using the CPO Ambler classification feature.
  2. Becton Dickinson & Company recalls BD¿Spinal Tray with BD¿Whitacre Needle 22 G x 3.5 in. and 5 moreFDAThis is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall…
  3. Becton Dickinson & Co. recalls BD GasPak EZ CO2 Pouch System and 1 moreFDABD identified an issue affecting certain lots of BD GasPak EZ CO2 and BD GasPak EZ Campy Gas Generating Pouch Systems, where gas generating sachets…
  4. Becton Dickinson & Company recalls mL Non-Sterile BD Luer-Lok Tip Syringe OnlyFDADuring the bulk packaging process, some Luer Slip syringes were incorrectly included in a batch designated for Luer Lok syringes.
  5. Becton Dickinson & Company recalls BD Luer Tip CapsFDABD has confirmed through internal testing that certain device trays failed routine biocompatibility testing.
  6. Becton Dickinson & Co. recalls BD Kiestra" Isolate Suspension Cuvette ArrayFDAThe 2D barcodes on these cuvettes contain formatting errors that prevent the analyzing system from correctly reading the expiration date. As a…
  7. Becton Dickinson & Co. recalls BD EpiCenter Microbiology Data Management System and 4 moreFDAProduct service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until…
  8. Becton Dickinson & Co. recalls BD COR System SoftwareFDAPotential for functionality issue that supports the Over labeling feature and could lead to possible false negative results.
  9. Becton Dickinson & Co. recalls BD BACTEC MGIT 960 PZA KitFDABD has received additional complaints of intermittent false resistance results for PZA during susceptibility testing of Mycobacterium tuberculosis…
  10. Becton Dickinson & Co. recalls BD BBL TB Stain Kit KFDAOne component of the kit was shipped with an expiration date of 30Sep2024 while the Certificate of Analysis (COA) indicated an expiration date of…
  11. Becton Dickinson & Co. recalls BD PCR Cartridges used with MAX SystemFDAPotential for signal drift in specific lots of BD PCR Cartridges.