Trokamed GmbH recalls Sheath
- Published
- 22 Jan 2025
- Units affected
- 4 products
- Agency
- FDA
- Country
- United States
The hazard
If the inflow speed is too quick, it could result in reduced outflow through the shaft due to blood, tissue fragments and kidney stone fragments, and quickly build up of intrarenal pressure.
Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Sheath, 18 Fr., 130 mm (REF: WA2PS18S), gastroenterology-urology endoscopic access overtube.
Sheath, 20 Fr., 160 mm (REF: WA2PS20L), gastroenterology-urology endoscopic access overtube.
Sheath, 20 Fr., 130 mm (REF: WA2PS20S), gastroenterology-urology endoscopic access overtube.
Sheath, 18 Fr., 160 mm (REF: WA2PS18L), gastroenterology-urology endoscopic access overtube.
- Brand / company
- Trokamed GmbH
- Codes / lots
- REF: WA2PS18S; UDI-DI: 04251303810865; Lot Numbers: 159825 | REF: WA2PS20L; UDI-DI: 04251303810919; Lot Numbers: 154236 | REF: WA2PS20S; UDI-DI: 04251303810926; Lot Numbers: 154125 | REF: WA2PS18L; UDI-DI: 04251303810872; Lot Numbers: 154214
- Distribution
- US (Domestic) distribution in state of PA and OUS (foreign) to country of Germany. .
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 22 Jan 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.