The hazard
Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.
Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Impella RP Flex with SmartAssist; Product Number: 1000323;
Impella RP with SmartAssist; Product Number: 0046-0035;
- Brand / company
- Abiomed, Inc.
- Codes / lots
- Product Number: 1000323; UDI-DI: 00813502012811; Serial Numbers: 551121, 551122, 551123, 551124, 551125, 551127, 551128, 551129, 537606, 537610, 537612, 537615, 537607, 537608, 537609, 537613, 537614, 548547, 548548, 548550, 548553, 548554, 548555, 548556, 548552, 548559, 548560, 548561, 548562, 548563, 548564, 548565, 549638, 549640, 549641, 549642, 549643, 548558, 546672, 546675, 546676, 546678, 546679, 546680, 547552, 547556, 547558, 547559, 547560, 546652, 546654, 546656, 546657, 546658, 546659, 547561, 549644, 549646, 549647, 546661, 547563, 547564, 547566, 547567, 547568, 547569, 547570, 547571, 546681, 547189, 547191, 547192, 547193, 547194, 547195, 548827, 548830, 548831, 548834, 548835, 548836, 543714, 543716, 543719, 543720, 543721, 543722, 547169, 547170, 547171, 547172, 547173, 547174, 547188, 548188, 548189, 548196, 545799, 545800, 545801, 545802, 545803, 545804, 548191, 548192, 548193, 548194, 545486, 545488, 545795, 545796, 545797, 529171, 529172, 529174, 529175, 529179, 545475, 545478, 545481, 545482, 545483, 545489, 545490, 545491, 545492, 545494, 539350, 539352, 539353, 539354, 539355, 543274, 543275, 543276, 543277, 543278, 543279, 543282, 529170, 543273, 543283,
- Distribution
- Domestic: Nationwide Distribution International: Canada
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 22 Jan 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.