CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Nova Biomedical Corporation recalls StatStrip Glucose Hospital Meter…

Nova Biomedical Corporation recalls StatStrip Glucose Hospital Meter System and 2 more

Read the official FDA notice

Published
1 Jan 2025
Units affected
3 products
Agency
FDA
Country
United States

The hazard

A software bug within the Gen 2 StatStrip Hospital Meter firmware leads to transmission of erroneous glucose patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current glucose test result to the DMS. The risk of this software defect is erroneously high or low glucose results being documented in a patient's medical record, which may lead to incorrect treatment.

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

StatStrip Glucose Hospital Meter System. Model/Catalog Number: 63685. Prescription use blood glucose meter for near-patient testing.

StatStrip Glucose Ketone (mmol/L) Hospital Meter System. Model/Catalog Number: 63910. Prescription use blood glucose ketone meter for near-patient testing

StatStrip Glucose Ketone Hospital Meter System. Model/Catalog Number: 63683. Prescription use blood glucose ketone meter for near-patient testing

Brand / company
Nova Biomedical Corporation
Codes / lots
Model No 63685. UDI-DI (01)10385480636858. Software Versions: v0.0.13.10 to v0.0.13.44 | Model No 63910. UDI-DI (01)1038548063910. Software Versions: v0.0.13.10 to v0.0.13.44. | Model No 63683. UDI-DI (01)38548063685. Software versions: v0.0.13.10 to v0.0.13.44.
Distribution
Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 1 Jan 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.