CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Moximed, Inc. recalls MISHA Knee System Implant Small

Read the official FDA notice

Published
1 Jan 2025
Units affected
115
Agency
FDA
Country
United States

The hazard

Knee implant may fracture due to a supplier-related issue with one component that has nonconforming geometry at the distal end of the absorber component that may result in a reduced wall thickness and microcracks, which may lead to new or worsening symptoms of discomfort, swelling / edema, pain or stiffness, which may necessitate removal of the device.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

MISHA Knee System Implant Small, Left, REF: 2-1001. MISHA Knee System Implant Small, Right, REF: 2-1002.

Brand / company
Moximed, Inc.
Codes / lots
REF/UDI-DI/Lot(Expiration): 2-1001/00856047005795/23032901(29-Mar-25), 23092102(21-Sep-25), 24032601(26-Mar-26). 2-1002/00856047005801/23032902(29-Mar-25), 23092503(25-Sep-25), 24040801(8-Apr-26).
Distribution
US Nationwide distribution in the states of CA, OR, NY, OH, TX, CO, NY, VA.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 1 Jan 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.