CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Maquet Cardiovascular, LLC recalls VasoView HemoPro 2 Endoscopic…

Maquet Cardiovascular, LLC recalls VasoView HemoPro 2 Endoscopic Vessel Harvesting System and 1 more

Read the official FDA notice

Published
15 Jan 2025
Units affected
2 products
Agency
FDA
Country
United States

The hazard

Potential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

VasoView HemoPro 2 Endoscopic Vessel Harvesting System. Model Numbers: VH-4000. Indicated for use in minimally invasive surgery allowing access for vessel harvesting.

Vasoview Hemopro 2 (w/Vasoshield) Endoscopic Vessel Harvesting System. Model Numbers: VH-4001. Indicated for use in minimally invasive surgery allowing access for vessel harvesting.

Brand / company
Maquet Cardiovascular, LLC
Codes / lots
Model Number: VH-4000. UDI-DI: 00607567700406. All unexpired Lots. | Model Number: VH-4001. UDI-DI: 00607567700451. All unexpired Lots.
Distribution
Nationwide distribution. International distribution to Australia, Austria, Bahrain, Bangladesh, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Guatemala, Hong Kong, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Kuwait, Malaysia, Mexico, Netherlands, Netherlands, New Zealand, Oman, Panama, Puerto Rico, Philippines, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 15 Jan 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.