The hazard
Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts and necessary clinical intervention could be delayed.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data.
- Brand / company
- Spacelabs Healthcare, Inc.
- Codes / lots
- Software Versions: 1.4.0 and 1.4.1. UDI-DI: 10841522107177, 10841522100246
- Distribution
- US: MN, TX, TN, SC, OK, NM, GA, PA, OH, IA, AL, CO, AR, FL, MO, ME, LA, IL, WA, MA, WI, OR, MS, ID, NY, KS, PR. OUS: Canada, Taiwan, Czech Republic, France, Greece, Italy, Kuwait, Philippines, Poland.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 15 Jan 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.