The hazard
Gas powered emergency resuscitator has a circuit flow controller manufactured with an undersized spring, which may cause resuscitator not reach the high and low ends of the PIP and PEEP pressure ranges, which may cause a loss of positive pressure, impacting effective ventilation of the patient.
Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Neo-Tee T-Piece Resuscitator, for pediatric emergency respiratory support, Part: 1050805, 1050808, 1050809, 1050810, 1050811, 1050814, 1050832, 1050839, 1050840, 1050841, 1050842
- Brand / company
- Mercury Enterprises, Inc. dba Mercury Medical
- Codes / lots
- Part/UDI-DI/Lot: 1050805/10641043508053, 30641043508057/2421450805, 2423550805, 2423650805, 2426750805; 1050808/10641043508084, 30641043508088/2426450808, 2426750808, 2426850808, 2429050808, 2429350808; 1050809/10641043508091, 30641043508095/2426450809, 2426750809, 2429050809, 2429150809, 2429350809, 2429750809, 2429850809, 2430050809, 2430250809; 1050810 /10641043508107, 30641043508101/2426750810, 2429350810, 2429750810, 2429850810, 2430250810; 1050811/10641043508114, 30641043508118/2426550811, 2426650811, 2426750811, 2429050811, 2429150811, 2429250811, 2429350811, 2429450811, 2429750811, 2429850811, 2429950811, 2430250811; 1050814/10641043508145, 30641043508149/2426450814, 2426750814, 2429050814, 2429750814, 2430250814; 1050832/10641043508329, 30641043508323/2429050832, 242985083; 1050839/10641043508398, 30641043508392/2430350839; 1050840/10641043508404, 30641043508408/2429050840, 2429750840; 1050841/10641043508411, 30641043508415/2429050841; 1050842/10641043508428, 30641043508422/2426750842
- Distribution
- Worldwide Distribution: US (nationwide) to states of: PA, FL, IN, WI, ME, SC, WY, ID, MT, TX, MI, LA, MS, AR, WV, MD, KY, NJ, WA, CA, GA, MO, MA, NC, NV, TN, VA, OK, ND, MN, AL, HI, UT, NY, AZ, IL, OH, IA, NE, CT, AK, OR and OUS (foreign) to countries of: Hungary, Switzerland, Hong Kong, Spain, Canada, Italy, United Kingdom, United Arab Emirates, Saudi Arabia, France, Netherlands
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 22 Jan 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.