CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Percussionaire Corporation recalls VDR4 Phasitron Breathing Circuit…

Percussionaire Corporation recalls VDR4 Phasitron Breathing Circuit Models that contain the venturi co…

Read the official FDA notice

Published
29 Jan 2025
Units affected
4,727 packs/cases
Agency
FDA
Country
United States

The hazard

Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation.

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK, Phasitron Kit, VDR, Single Patient, 5pk -A51001, VDR4 Control Unit Tester -PRT 991, VDR W/SWIVEL T SINGLE PATIENT PHASITRON -PRT 992, VDR4 HUMIDIFICATION ADAPTER KIT, CASE OF 10 -PRT 993, VDR4 Humidification Kit with Cross Tee, Case of 10 The Phasitron breathing circuit kit is intended to be used for continuous ventilation intended for controlled ventilation of patients

Brand / company
Percussionaire Corporation
Codes / lots
-Phasitron Kit, VDR, Single Patient, 5pk Model Number - A50094-D-5PK UDI-DI code: 00849436000259 Lot Codes: WO04294 WO04424 WO04764 WO05070 WO05186 WO05460 WO05685 WO05910 WO06388 WO06576 WO06883 WO07095 WO07196 WO07317 WO07405 WO07450 WO07696 - Control Unit Tester Model Number: A51001 VDR4 Lot Codes: WO04750 WO06701 - VDR W/Swivel T Single Patient Phasitron Model Number: PRT 991 Lot Code: WO04733 - VDR4 Humidification Adapter Kit, Case of 10 Model Number: PRT 992 Lot Code: WO045667 - VDR4 Humidification Kit with Cross Tee, Case of 10 Model Number: PRT 993 Lot Codes: WO04745 WO04832 WO04893 WO04893 WO05309 WO06523 WO07080 WO07283
Distribution
Worldwide - U.S. Nationwide distribution in the states of AZ, FL, IL, MS, MO, NE, NJ, NY, NC, RI, TX, UT, and WV. The countries of Belgium, Canada, France, Hong Kong, Netherlands, Russia, Switzerland and Turkey.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 29 Jan 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.