SIN SISTEMA DE IMPLANTE NACIONAL S.A recalls Implant Epikut S Plus CM…

SIN SISTEMA DE IMPLANTE NACIONAL S.A recalls Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm

Read the official FDA notice

Published
22 Jan 2025
Units affected
123 (US 98, OUS 25)
Agency
FDA
Country
United States

The hazard

Dental implant box incorrectly labels the diameter as 4,5 X 18,0 mm, but the internal label correctly labels the diameter as 4,0 X 18,0 mm; if the dentist drills the bone with the larger diameter stated on the box, the smaller diameter implant, will become loose and will not fit, which may necessitate an additional surgery.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N

Brand / company
SIN SISTEMA DE IMPLANTE NACIONAL S.A
Codes / lots
UDI-DI: 07899995273437, Lot: X010469757, Expiration: 2028/01
Distribution
US distribution to states of: KY, CO, CA, NY, ID, NJ, FL, TX, KY, IL, OH; AND OUS (foreign) country of Brazil.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 22 Jan 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.