CheckMyRecalls

Nobel Biocare recalls products

Published
12 Feb 2025
Units affected
1,558 units (90 units - U.S.)
Agency
FDA
Country
United States

The hazard

Due to manufacturing issue (unintended by-product of the injection molding process) a sharp pin that may puncture the blister packaging and breaching the sterile barrier of the finished product.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

. Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 TiUltra TCC RP 4.0x11mm . Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Nobel Biocare N1 TiUltra TCC implants are intended for use as an endosseous dental implant in the maxilla or mandible for anchoring or supporting dental protheses to restore chewing function

Brand / company
Nobel Biocare
Codes / lots
Nobel Biocare N1 TiUltra TCC RP 4.0x 9mm Catalog Number: 300861 UDI-DI code: 07332747161731 Lot Numbers: 12253672 12253673 Nobel Biocare N1 TiUltra TCC RP 4.0x11mm Catalog Number: 300862 UDI-DI code: 07332747161748 Lot Numbers: 12253448 12254209 Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Catalog Number: 300863 UDI-DI code: 07332747161755 Lot Numbers: 12253620 12254213
Distribution
U.S. Nationwide distribution in the states of CA, FL, IL, MA, NJ, NY, OH, PA, UT, VA and WA.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 12 Feb 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.