CareFusion 303, Inc. recalls BD Alaris Pump Module REF 8100 The BD…
CareFusion 303, Inc. recalls BD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails…
- Published
- 5 Mar 2025
- Units affected
- 242,098 systems
- Agency
- FDA
- Country
- United States
The hazard
Due to User Manual/Direction for Use incorrectly instructing user to leave the safety clamp in the open position when reloading an infusion set. This issue potentially could result in an accidental over infusion if the roller clamp is left open.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
BD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration
- Brand / company
- CareFusion 303, Inc.
- Codes / lots
- Catalog Number: 8100 UDI-DI codes: 10885403810046 10885403517723 Software Version: 12.1.2 Serial/Lot Numbers: Pending
- Distribution
- U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and Guantanamo Bay, Cuba O.U.S.: Canada
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 5 Mar 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.