Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS)

Read the official FDA notice

Published
12 Feb 2025
Units affected
23 systems
Agency
FDA
Country
United States

The hazard

Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty.

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.

Brand / company
Fresenius Kabi USA, LLC
Codes / lots
UDI-DI: 00811505030122. Software 5.9.2 and prior
Distribution
US distribution to CA, CO, GA, ID, MI, MS, NE, NJ, NV, OK, TX, UT, VA, WA, WI.
Recall status
Ongoing