ROi CPS LLC recalls regard Item Number

Published
12 Feb 2025
Units affected
414 kits
Agency
FDA
Country
United States

The hazard

There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.

Brand / company
ROi CPS LLC
Codes / lots
Lot number 104006, Exp. 9/30/2026, UDI (01)10194717119074(17)260930(10)104006; and Lot number 103486, Exp. 8/31/2026, UDI (01)10194717119074(17)260831(10)103486.
Distribution
Distribution was made to MO. There was no government/military/foreign distribution.
Recall status
Ongoing