The hazard
Potential failure to alarm for "No Device Data" INOP. The device to alarm after a Hamilton ventilator is disconnected from the EC10 Module or IntelliBridge I/O board.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
IntelliVue Patient Monitor MX400. Model Number: 866060
IntelliVue Patient Monitor MX450. Model Number: 866062
IntelliVue Patient Monitor MX550. Model Number: 866066
IntelliVue Patient Monitor MX500. Model Number: 866064
IntelliBridge EC10 Module. Model Number: 865115
- Brand / company
- Philips North America Llc
- Codes / lots
- Model Number: 866060. UDI-DI Number: (01)00884838038752(21). Serial Numbers: DE35174105, DE35174084, DE35174083, DE35174082, DE35174080, DE35107369, DE35108320, DE35160410, DE671M4116, DE7770DA9W, DE35114616, DE35114618, DE671E8921, DE671E8917, DE7580AH50, DE7580AHEH, DE35123130, DE671V9297, DE671V9296, DE671V9298, DE35160989, DE35160994, DE35160988, DE35161272, DE35161273, DE35161274, DE7580B537, DE7580B52H. | Model Number: 866062. UDI-DI: (01)00884838038769(21); Serial Numbers: DE35169761, DE671V3903, DE671V3885, DE671V7898, DE7770DC1L. UDI-DI: 884838045859; Serial Numbers: DE32303399, DE32302479, DE32302480, DE32306956, DE35131294, DE35114613, DE35114612, DE671T9641, DE7770C86L, DE7770C86K, DE7770C86J, DE7770C86H, DE7770C86G, DE35120431, DE671K3731, DE671K3729, DE671K3706, DE671K3796, DE671K3816, DE671K3737, DE671K3773, DE671L0436, DE671L0462, DE671L0406, DE671L5214, DE671L5207, DE671L5206, DE671L5181, DE671L5169, DE671M3993, DE671M4048, DE671M4036, DE671M4037, DE671M4053, DE671M7009, DE671M7012, DE671M9968, DE671M9972, DE671M9963, DE671N0009, DE671N0010, DE671N0092, DE671N0008, DE671N0001, DE671N0002, DE671M9985, DE671M9999, DE671N0000, DE671P0411, DE671P1094, DE671P2457, DE671
- Distribution
- US distribution to Florida, North Carolina, and Oregan. International distribution to Canada, Estonia, France, Germany, Japan, Netherlands, Spain, Switzerland, United Kingdom.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 12 Feb 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.