CheckMyRecalls

Microbiologics Inc recalls Microbiologics Multi-Strain Human…

Microbiologics Inc recalls Microbiologics Multi-Strain Human Papillomavirus (HPV) Control Pane… and 1 more

Published
19 Feb 2025
Units affected
2 products
Agency
FDA
Country
United States

The hazard

Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Microbiologics Multi-Strain Human Papillomavirus (HPV) Control Panel (Inactivated Pellet), REF 8234; in-vitro diagnostic test

Microbiologics Multi-Strain Human Papillomavirus (HPV) Verification Panel (Inactivated Pellet), REF 8235; in-vitro diagnostic test

Brand / company
Microbiologics Inc
Codes / lots
UDI/DI 70845357043077, Lot Number 8234-22 | UDI/DI 70845357043077, Lot Numbers: 8235-32, 8235-33
Distribution
Worldwide distribution - US Nationwide and the countries of Netherlands, Sweden, UK, France, Australia, New Zealand.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 19 Feb 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.