The hazard
Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Microbiologics Multi-Strain Human Papillomavirus (HPV) Control Panel (Inactivated Pellet), REF 8234; in-vitro diagnostic test
Microbiologics Multi-Strain Human Papillomavirus (HPV) Verification Panel (Inactivated Pellet), REF 8235; in-vitro diagnostic test
- Brand / company
- Microbiologics Inc
- Codes / lots
- UDI/DI 70845357043077, Lot Number 8234-22 | UDI/DI 70845357043077, Lot Numbers: 8235-32, 8235-33
- Distribution
- Worldwide distribution - US Nationwide and the countries of Netherlands, Sweden, UK, France, Australia, New Zealand.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 19 Feb 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.