American Contract Systems Inc recalls CRANI PACK
- Published
- 26 Mar 2025
- Units affected
- 91 units
- Agency
- FDA
- Country
- United States
The hazard
During an internal investigation, ACS was unable to confirm that the CODMAN component achieved the required Sterility Assurance Level and therefore is not suitable for sterilization according to ACS procedures.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
CRANI PACK, Model/Catalog Number WECR41D, medical procedure kit.
- Brand / company
- American Contract Systems Inc
- Codes / lots
- Lot Code: UDI: 191072216984, Lot Cocdes: 71-052536, Exp: 10/30/2025; 71-051491, Exp: 07/16/2025; LOT: 71-051609, Exp: 07/30/2025; 71-051767, Exp: 08/14/2025; 71-053065, Exp: 12/10/2025
- Distribution
- US: MI
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 26 Mar 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.