The hazard
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
- Brand / company
- IHB OPERATIONS B.V.
- Codes / lots
- UDI/DI 10841736116071, All Serial Numbers
- Distribution
- US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 30 Apr 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.