Globus Medical, Inc. recalls MAGEC 2 Rod

Read the official FDA notice

Published
30 Apr 2025
Units affected
1 unit
Agency
FDA
Country
United States

The hazard

Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

MAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;

Brand / company
Globus Medical, Inc.
Codes / lots
Version/Model: MC2-5090S; UDI-DI (with lot nuber): (01)00887517834041(11)250311(17)300311(10)GB1607ET
Distribution
US Nationwide distribution in the state of AR.
Recall status
Ongoing