The hazard
Balloon extension lumens are not printed correctly. Specifically, the inner balloon inflation hub was labeled as the outer balloon, and the outer balloon inflation hub was labeled as the inner balloon.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
BIB Balloon in Balloon Dilatation Catheter. Model Number: 420
- Brand / company
- Numed Inc
- Codes / lots
- Model Number 420. UDI-DI: 04046964899768. Lot Numbers: BIB-16972 (Expiration date: 11/30/2028), BIB-16972-1 (Expiration date: 12/31/2028). Catheter Product Code: BB005.
- Distribution
- US distribution to Pennsylvania.
- Recall status
- Terminated
Source: U.S. Food and Drug Administration, published 21 May 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.