CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Silex Medical, LLC recalls Signature Laparoscopic Instruments

Read the official FDA notice

Published
11 Jun 2025
Units affected
8 products
Agency
FDA
Country
United States

The hazard

Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Signature Laparoscopic Instruments, Bullet Tip Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5863030.

Signature Laparoscopic Instruments, Debakey Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5862100.

Signature Laparoscopic Instruments, Sharp Tooth Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5863020.

Signature Laparoscopic Instruments, Dolphin Nose Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5860020.

Signature Laparoscopic Instruments, Atraumatic Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5864000.

Signature Laparoscopic Instruments, Alligator Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5863050.

Signature Laparoscopic Instruments, Allis Clamp Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5862030.

Signature Laparoscopic Instruments, 20mm Atraumatic Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5881033.

Brand / company
Silex Medical, LLC
Codes / lots
Model/REF Number: 5863030. UDI-DI: B33158630300. Lot Number: M28185. Serial Numbers: 0446, 0447, 0448, 0449, 0450, 0451. | Model/REF Number: 5862100. UDI-DI: B33158621000. Lot Number: M28183. Serial Numbers: 1932, 1933, 1934, 1935, 1936, 1937, 1938, 1939. | Model/REF Number: 5863020. UDI-DI: B33158630200. Lot Number: M28184. Serial Numbers: 2222, 2223, 2224, 2225, 2226, 2227, 2228, 2229, 2230, 2231, 2232, 2233, 2234, 2235, 2236, 2237, 2238, 2239, 2240, 2241, 2242, 2243, 2244, 2245, 2246, 2247, 2248, 2249, 2250, 2251, 2252, 2253 | Model/REF Number: 5860020. UDI-DI: B33158600200. Lot Number: M28181. Serial Numbers: 0691, 0692. | Model/REF Number: 5864000. UDI-DI: B33158640000. Lot Number: M28187. Serial Numbers: 9565, 9566, 9567, 9568, 9569, 9570, 9571, 9572, 9573, 9574, 9575, 9576, 9577, 9578, 9579, 9580, 9581, 9582, 9583, 9584, 9585, 9586, 9587, 9588, 9589, 9590, 9591, 9592, 9593, 9594, 9595, 9596, 9597, 9598, 9599, 9600, 9601, 9602, 9603, 9604, 9605, 9606, 9607, 9608, 9609, 9610, 9611, 9612, 9613, 9614, 9615, 9616, 9617, 9618, 9619, 9620, 9621, 9622, 9623, 9624, 9625, 9626, 9627, 9628, 9629, 9630, 9631, 9632, 9633, 9634, 9635, 9636, 9637, 9638, 9639, 9640, 9641, 9642, 9643, 9644,
Distribution
US Nationwide distribution in the states of Florida, North Dakota, Illinois, California, Texas, Virginia, Michigan, Connecticut, Oregon, Iowa, Mississippi, New Jersey, Idaho, West Virginia, Hawaii.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 11 Jun 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.