The hazard
Potential for shorter clotting times in clotting tests which have a dilution, leading to possible overestimate of around 1%
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
STA - Owren Koller. Model/Catalog Number: 00360 STA - Owren-Koller is a buffer solution intended for use as a diluent for reagents and human plasmas in coagulation tests performed with analyzers of the STA-R, STA Compact, STA Satellite and ST art families.
- Brand / company
- Diagnostica Stago, Inc.
- Codes / lots
- UDI: (01)13607450003601(11)240331(17)260331(10)270902(241)00360. Lot Number: 270902.
- Distribution
- The affected lot was distributed to customers in various US states in October and November 2024. The lot was also distributed in other countries including Canada. The lot was provided to two Distributors in US Territories: MedPharm Corp (Guam) and Isla Lab Products (Puerto Rico); MedPharm will be notified by Stago's Asia-Pacific business unit in the coming days and is not part of this report.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 25 Jun 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.