The hazard
GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound system
GE Healthcare Versana Ultrasound Systems, sold under the following names and Model/Catalog Numbers: 1) Versana Premier R3 VS, Model/Catalog Number 5938558; 2) Versana Premier R3 VA, Model/Catalog Number 5938559; 3) Versana Premier Lotus 4PP, Model/Catalog Number 5938560; 4) Versana Premier Lotus 5PP, Model/Catalog Number 5938561; 5) Versana Premier R3 VA Vet, Model/Catalog Number 5946387; 6) Versana Premier R3 VS Vet, Model/Catalog Number 5946481. 7) Versana Premier R3 Expert, Model/Catalog Number 5946950; 8) Versana Premier R3, Model/Catalog Number 5946951; 9) Versana Premier R3 Pro, Model/Catalog Number 5946952; 10) Versana Premier R3 Elite, Model/Catalog Number 5946953; 11) Versana Premier R3 Plus, Model/Catalog Number 5946954; 12) Versana Premier R3 Ultra, Model/Catalog Number 5946955; 13) Versana Premier R3 Max, Model/Catalog Number 5946956; 14) Versana Premier R3 Super, Model/Catalog Number 5946957; 15) Versana Premier R3 VS India, Model/Catalog Number 5948398; 16) Versana Premier R3 VA India, Model/Catalog Number 5948399; diagnostic ultrasound system
- Brand / company
- GE Medical Systems China Co., Ltd.
- Codes / lots
- UDI/DI 00195278722430, Lot/Serial Numbers: 6310028WX0, 6310008WX0, 6310041WX0, 6310017WX0, 6310031WX0, 6310044WX0, 6310024WX0, 6310038WX0, 6310040WX0, 6310019WX0, 6310018WX0, 6310023WX0, 6310033WX0, 6310012WX0, 6310026WX0, 6310020WX0, 6310032WX0, 6310014WX0, 6310021WX0, 6310025WX0, 6310010WX0, 6310027WX0, 6310037WX0, 6310009WX0, 6310007WX0, 6310035WX0, 6310016WX0, 6310039WX0, 6310036WX0, 6310043WX0, 6310011WX0, 6310015WX0, 6310013WX0, 6310029WX0. | Model/Catalog Number 5938558, UDI/DI (01)00195278690388(11)241127(21)6302036WX0, Lot/Serial Number 6302036WX0; Model/Catalog Number 5938558, UDI/DI (01)00195278690388(11)241127(21)6302045WX0, Lot/Serial Number 6302045WX0; Model/Catalog Number 5938558, UDI/DI (01)00195278690388(11)241127(21)6302035WX0, Lot/Serial Number 6302035WX0; Model/Catalog Number 5938558, UDI/DI (01)00195278690388(11)241127(21)6302046WX0, Lot/Serial Number 6302046WX0; Model/Catalog Number 5938558, UDI/DI (01)00195278690388(11)241127(21)6302033WX0, Lot/Serial Number 6302033WX0; Model/Catalog Number 5938558, UDI/DI (01)00195278690388(11)241127(21)6302034WX0, Lot/Serial Number 6302034WX0; Model/Catalog Number 5938558, UDI/DI (01)00195278690388(11)241127(21)6302041WX0,
- Distribution
- Worldwide distribution.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 2 Jul 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.