CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Maquet Critical Care AB recalls Servo-u Ventilator System and 2 more

Read the official FDA notice

Published
9 Jul 2025
Units affected
3 products
Agency
FDA
Country
United States

The hazard

Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Servo-u Ventilator System. Model Number: 6688600.

Servo-n Ventilator System. Model Number: 6694800.

Servo-u MR Ventilator System. Model Number: 6888800.

Brand / company
Maquet Critical Care AB
Codes / lots
Model Number: 6688600. UDI-DI: 07325710001042. All software versions. | Model Number: 6694800. UDI-DI: 07325710001110. All software versions. | Model Number: 6888800. UDI-DI: 07325710010518. All software versions.
Distribution
US Nationwide distribution and Puerto Rico. No international distribution.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 9 Jul 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.