CheckMyRecalls

Waismed Ltd. recalls Brand Name

Published
23 Jul 2025
Units affected
2 products
Agency
FDA
Country
United States

The hazard

Due to manufacturing error, Intraosseous device built-in stabilizer may not be released properly after deployment and insertion.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Brand Name: NIO+ Product Name: NIO Intraosseous Device Adult Model/Catalog Number: 105000908 Software Version: N/A Product Description: NIO devices are a spring-based, automatic intraosseous access device supplied Sterile and single use. Indicated for Intraosseous access to the Proximal humerus and Proximal Tibia in adult patients, in emergent situations. Component: N/A

Brand Name: NIO-A Product Name: NIO Intraosseous Device Adult Model/Catalog Number: 105000603 Software Version: N/A Product Description: NIO devices are a spring-based, automatic intra-osseous access device supplied Sterile and single use. Indicated for Intra-osseous access to the Proximal humerus and Proximal Tibia in adult patients, in emergent situations. Component: N/A

Brand / company
Waismed Ltd.
Codes / lots
Lot Code: UDI: 07290008325073 Lot number: 2440044 | UDI: 0 7290008325059 Lot number: 2440043, 2440045, 2540046, 2540047
Distribution
Worldwide distribution: US (nationwide) to state of: NC and OUS (international) countries of: Slovenia, Australia, Brazil, Estonia, South Africa, U. Arab Emirate, Greece, Panama, Singapore, Italy, Germany, Portugal, Spain, Ukraine, Lithuania, Chile, Poland, Germany, New Zealand, Montenegro, Switzerland, Argentina, Poland, Austria, Sweden, United Kingdom, Italy, France, Romania, Netherlands, Croatia, Lithuania, New Zealand, Japan, Cyprus, and Israel.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 23 Jul 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.