The hazard
During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient and may cause thermal injury to the patient s mucous membrane.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Pentax Medical Video Processor; Model Number: EPK-i8020c;
- Brand / company
- Pentax of America Inc
- Codes / lots
- Model Number: EPK-i8020c; UDI-DI: 04961333247974; Serial Numbers: B0023Z1657, D0023Z0245, D0023Z0206, D0023Z0125, D0023Z0123, D0023Z0119, D0023Z0118, D0023Z0117, D0023Z0114, D0023Z0112, D0023Z0109, D0023Z0108, D0023Z0011, C0023Z0087, C0023Z0085, C0023Z0074, C0023Z0072, C0023Z0021, B0023Z1712, B0023Z1711, B0023Z1670, B0023Z1665, B0023Z1659, B0023Z1653, B0023Z1628, b0023z1501, B0023Z1195, B0023Z1707, C0023Z0084, B0023Z1732, B0023Z1669, D0023Z0137, D0023Z0141, D0158Z0049, D0158Z0033, E0023Z0022, D0158Z0053, D0158Z0019, E0023Z0026, E0023Z0008, D0023Z0248, D0023Z0241, D0023Z0222, D0023Z0155, B0023Z1276, C0023Z0082, D0023Z0233, B0023Z1461, D0023Z0139, D0023Z0140, D0023Z0122, B0023Z1709, E0023Z0020, E0023Z0009, E0023Z0002, D0023Z0142, E0023Z0024, E0023Z0018, E0023Z0013, E0023Z0004, C0023Z0095, C0023Z0078, E0023Z0007, E0158Z0057, E0158Z0035, B0023Z1735, B0023Z1717, B0023Z1648, B0023Z1639, B0023Z1631, C0158Z0047, C0158Z0042, B0023Z1481, B0023Z1478, B0023Z1451, B0023Z1335, B0023Z1193, B0023Z1288, B0023Z1258, B0023Z1296, B0023Z1290, B0023Z1286, B0023Z1280, B0023Z1279, B0023Z1256, B0023Z1250, B0023Z1245, B0023Z1676, B0023Z1573, B0023Z1565, B0023Z1656, B0023Z1694, D0158Z0055, B0023Z1577, B0023Z
- Distribution
- US Distribution to states of: AZ, CA, DC, FL IA, ID, IL, KS, MA, MD, MI, MO, NC, NJ, NM, NY, OK, PA, TX, WA.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 27 Aug 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.