The hazard
The firm issued a field safety notice after becoming aware of three lots of products not having quarterly dose audits being completed on time in the first quarter of 2024. Routine lot by lot testing was completed properly however the quarterly dose audit (sampling/testing to monitor the gamma sterilization process and confirm the validated parameters remain effective to assure sterility and bioburden reduction per requirements) was not completed on time.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Bridge to Life EasiSlush, (Sodium Chloride Solution for Sterile Slush Preparation) (0.9% Sodium Chloride Irrigation, USP), Part Number REF BTLE-1250
- Brand / company
- Bridge to Life Ltd
- Codes / lots
- UDI 10860001991205-240206-260206-240372708 (case), 00860001991208-240206-260206-240372708 (bag), Lot Number 240372708; UDI 10860001991205-240212-260212-240402836 (case), 00860001991208-240212-260212-240402836 (bag), Lot Number 240402836; UDI 10860001991205-240214-260214-240402837 (case), 00860001991208-240214-260214-240402837 (bag), Lot Number 240402837.
- Distribution
- United States - New York, Maryland, Alabama, North Carolina, South Carolina, Ohio, Florida, Wisconsin, Utah, Tennessee, New Jersey, Georgia, Nebraska, Arizona, Minnesota, California, and Canada
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 30 Jul 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.