B BRAUN MEDICAL INC recalls Microbore Extension Set with 0.2 Micron…

B BRAUN MEDICAL INC recalls Microbore Extension Set with 0.2 Micron Filter

Read the official FDA notice

Published
13 Aug 2025
Units affected
34,825 units US
Agency
FDA
Country
United States

The hazard

Label is incorrect. The product label indicates that the device contains an air eliminating filter. However, the filter utilized on this extension set is not indicated for removal of air and does not feature an air vent, which is common to air eliminating filters.

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Microbore Extension Set with 0.2 Micron Filter. Catalog Number: V6215. for use with IV tubing.

Brand / company
B BRAUN MEDICAL INC
Codes / lots
Catalog Number: V6215. Primary UDI-DI 04046964189173, Unit of Use UDI-DI: 04046964189166. Lot Number (Expiration Date): 0061742452 (30JUN2025), 0061747379 (31AUG2025 0, 0061767411 (31JAN2026), 0061780914 (30APR2026), 0061803499 (30SEP2026), 0061806173 (30NOV2026), 0061822333 (30APR2027), 0061836578 (30APR2027), 0061849109 (31JUL2027), 0061850836 (31AUG2027), 0061899902 (30SEP2028), 0061936368 (31MAR2029), 0061936119 (31MAR2029), 0061936369 (31MAR2029), 0061940584 (30APR2029).
Distribution
US distribution to AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, MA, MD, ME, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, VA, WA, WY. International distribution to Canada.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 13 Aug 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.