Capnia Inc recalls Precision Sampling Set
- Published
- 6 Aug 2025
- Units affected
- 2160
- Agency
- FDA
- Country
- United States
The hazard
Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor
- Brand / company
- Capnia Inc
- Codes / lots
- Monitor UDI-DI: 00858481006017. Lot (C20206): 231220
- Distribution
- US Nationwide distribution in the states of PA, UT, TX, FL, MA, IA.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 6 Aug 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.