The hazard
Low deaeration chamber level at post priming tests may lead to PrisMax System alarm T2309: Air Detected in Prime resulting in multiple additional priming cycles
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
PrisMax V2-US, Hemodialysis Delivery System, Product Codes: 955626, 955558, 955627, 955701
TherMax Blood Warmer Unit-US, PrisMax Accessory, Product Codes: 955630, 955515, 955702, 955631
PrisMax V3 control Unit-US, Hemodialysis Delivery System, Product Code 955724
- Brand / company
- VANTIVE US HEALTHCARE LLC
- Codes / lots
- UDI 00085412639499, All Serial Numbers | UDI 07332414124731, All serial numbers | UDI 07332414126018, All serial numbers
- Distribution
- Worldwide
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 3 Sept 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.