The hazard
FilmArray NGDS Warrior Panel (NGDS-ASY-0007) has an increased risk of internal control failures when testing positive blood culture, which may lead to delayed results and additional diagnostic workup.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
FilmArray NGDS Warrior Panel NGDS-ASY-0007 is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the FilmArray 2.0 system.
- Brand / company
- Biofire Defense
- Codes / lots
- UDI-DI: 00851458005136. Lot(Expiration): 324324D(2026-02-19), D241022(2026-01-22), 319524D(2025-11-20), 315424D(2025-10-09), 315224D(2025-10-03), 315324D(2025-10-08)
- Distribution
- US Nationwide distribution in the states of AL, AZ, UT, CA, TX, DC , VA, FL, MD, OH, HI, NY, KY, LA, MA, MI, MN, MO, NC, NH, RI, NV, TN, IA, WA.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 1 Oct 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.