The hazard
When fenestrated analysis with clock measurements is started in diagnostic bioimaging software (intended to measure/visualize cardiovascular structures) and the 12h position of a single clock is changed, other clock measurements are not updated relative to the new 12h position which may cause stent graft fenestrations at incorrect position, which may lead to blood flow disruption and tissue damage
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
3mensio Workstation (Vascular Fenestrated) software
- Brand / company
- PIE Medical Imaging B.V.
- Codes / lots
- Software Version/UDI/Software Release Date: 10.6/(01)08056304455505(11)240325(8012)10.6/05-Apr-24, 10.7/(01)08056304455925(11)250423(8012)10.7/29-Apr-25, Service Packs/UDI: 10.6 SP1/(01)08056304455505(11)240411(8012)10.6 SP1/23-Apr-24, 10.6 SP2/(01)08056304455505(11)240610(8012)10.6 SP2/17-Jun-24, 10.6 SP3/(01)08056304455505(11)240926(8012)10.6 SP3/03-Oct-24, 10.6 SP4/(01)08056304455505(11)250109(8012)10.6 SP4/20-Jan-25, 10.7 SP1/(01)08056304455925(11)250523(8012)10.7 SP1/28-May-25
- Distribution
- US Nationwide distribution in the states of MN, NJ, GA, NY, OH, KS, MA, CA, TX, MS, NE, NC.
- Recall status
- Ongoing