CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Tandem Diabetes Care, Inc. recalls t:slim X2 insulin pump with…

Tandem Diabetes Care, Inc. recalls t:slim X2 insulin pump with Interoperable Technology

Read the official FDA notice

Published
29 Oct 2025
Units affected
122,958
Agency
FDA
Country
United States

The hazard

A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide when using Dexcom G7 sensor; Tandem Mobi insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014801, 1014811, 1014813, and User Guide when using Dexcom G7 sensor

Brand / company
Tandem Diabetes Care, Inc.
Codes / lots
t:slim X2, UDI-DI: 00389152407012, 00389152407111, 00389152407210, 00389152407319 with software versions 7.9.0.1 and 7.10.1, User Guide versions not including and prior to: AW-1014562 Revision D; Tandem Mobi, UDI-DI: 00389152480114, 00389152481111, 00389152481319 with software version 7.9.0.1, User Guide versions not including and prior to: AW-1014938 Revision C
Distribution
US nationwide distribution including Puerto Rico.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 29 Oct 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.