CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Zap Surgical Systems recalls ZAP-X Radiosurgery System

Read the official FDA notice

Published
8 Oct 2025
Units affected
17
Agency
FDA
Country
United States

The hazard

If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

ZAP-X Radiosurgery System, Model: 300150

Brand / company
Zap Surgical Systems
Codes / lots
UDI-DI: 00860183001504, Serial Numbers: ZUC17153, ZUC20156, ZUC20157, ZUC20159, ZUC21162, ZUC21164, ZUC22166, ZUC22169, ZUC22170, ZUC22171, ZUC22173, ZUC22174, ZUC23176, ZUC23177, ZUC24179, ZUC24180, ZUC24181. Software version: TDS V1.10.1
Distribution
Worldwide - US Nationwide distribution in the states of FL, CO, NJ, NY and the countries of Germany, France, Spain, Poland, Republic of Korea, T¿rkiye, Paraguay.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 8 Oct 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.