CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20…

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20 Biplane Product Codes and 12 more

Read the official FDA notice

Published
1 Oct 2025
Units affected
13 products
Agency
FDA
Country
United States

The hazard

Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Allura Xper FD20 Biplane Product Codes: (1)722013, (2) 722008; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Allura Xper FD20/15; Product Code: 722058; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Allura Xper FD10/10 Product Codes: (1)722011, (2) 722027, (3) 722005; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Allura Xper FD20/15 OR Table; Product Code: 722059; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Allura Xper FD10 OR Table Product Codes: (1)722022, (2) 722033; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Allura CV20; Product Code: 722031; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Allura Xper FD10 Product Codes: (1)722010, (2) 722026, (3) 722003; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Allura Xper FD20 OR Table Product Codes: (1)722023, (2) 722035, (3) 722015; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Allura Xper FD20/20 OR Table; Product Code: 722039; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Allura Xper FD20 Biplane OR Table Product Codes: (1)722020, (2) 722025; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Allura Xper FD20 Product Codes: (1)722012, (2) 722028, (3) 722006; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Allura Xper FD20/20; Product Code: 722038; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Brand / company
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Codes / lots
Product Codes: (1)722013, (2) 722008; UDI-DI: (1) 00884838059061, (2) N/A; Serial numbers: (1) 505, 313, 34, 153, 410, 360, 290, 241, 404, 126, 244, 106, 346, 383, 461, 307, 442, 435, 293, 289, 267, 214, 402, 86, 603, 149, 424, 140, 254, 255, 266, 518, 519, 303, 115, 150, 129, 496, 441, 124, 151, 550, 575, 562, 178, 330, 309, 601, 308, 264, 268, 141, 175, 85, 216, 422, 174, 228, 511, 152, 561, 262, 626, 497, 170, 379, 125, 564, 426, 495, 500, 515, 105, 428, 427, 531, 375, 316, 278, 347, 462, 249, 246, 639, 304, 147, 607, 567, 168, 171, 265, 332, 295, 589, 444, 447, 400, 384, 538, 117, 620, 624, 623, 215, 416, 415, 47, 335, 619, 598, 361, 358, 549, 133, 476, 315, 616, 630, 417, 490, 528, 525, 247, 618, 372, 613, 478, 544, 585, 608, 296, 570, 547, 138, 318, 345, 139, 473, 388, 434, 502, 586, 334, 411, 501, 458, 261, 429, 353, 395, 128, 348, 503, 557, 493, 563, 250, 595, 406, 355, 57, 394, 535, 569, 522, 465, 533, 412, 95, 344, 354, 425, 542, 207, 555, 524, 574, 366, 527, 325, 470, 440, 144, 403, 448, 134, 380, 487, 374, 256, 405, 367, 455, 302, 300, 299, 239, 506, 243, 324, 311, 70, 161, 449, 321, 176, 181, 554, 167, 130, 378, 158, 237, 453, 165, 532, 223, 343, 291, 370, 193, 163, 52
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Finland, France, FrenchPolynesia, Georgia, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Lao, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 1 Oct 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.