CheckMyRecalls

ImaCor Inc. recalls ClariTEE Miniaturized TEE probe

Published
15 Oct 2025
Units affected
45 units
Agency
FDA
Country
United States

The hazard

Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultrasound, ImaCor model ZHH.

Brand / company
ImaCor Inc.
Codes / lots
Model CLT-010; UDI-DI 00861589000108; Lot number: 20261209.
Distribution
US Nationwide distribution in the states of South Carolina, New Jersey, Pennsylvania, Florida, and Georgia.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 15 Oct 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.