CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Insulet Corporation recalls The failure occurs if a user on the…

Insulet Corporation recalls The failure occurs if a user on the Omnipod 5 iOS application selec…

Read the official FDA notice

Published
22 Oct 2025
Units affected
174,013 units
Agency
FDA
Country
United States

The hazard

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.

Brand / company
Insulet Corporation
Codes / lots
Omnipod 5 App Software. UDI: 10385083000459. All software versions 2.X.X
Distribution
US Nationwide distribution.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 22 Oct 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.