CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

BigTree Sales Inc. recalls The product is a handheld ultraviolet-C…

BigTree Sales Inc. recalls The product is a handheld ultraviolet-C germicidal wand and contain…

Read the official FDA notice

Published
24 Sept 2025
Units affected
N/A
Agency
FDA
Country
United States

The hazard

The subject product emits unsafe level of UVC radiation (i.e., the product reaches the exposure limit for Risk Group 3 (high risk) in about 2.2 minutes). The product has a radiation safety defect, in that these radiation emissions are unnecessary to the accomplishment of the product s primary purpose of disinfecting objects and surfaces, and create a risk of skin and/or eye injuries to user and nearby persons.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

The product is a handheld ultraviolet-C germicidal wand and contains a narrow-diameter, tubular lamps with a peak emission wavelength in the 254 nm range.

Brand / company
BigTree Sales Inc.
Codes / lots
SKU BT-GERMWAND-LRG-WHT and Testing: September 8-10, 2020
Distribution
U.S.A
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 24 Sept 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.