Wipro GE Healthcare Private Ltd. recalls GE Healthcare Carescape…

Wipro GE Healthcare Private Ltd. recalls GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop wi…

Read the official FDA notice

Published
29 Oct 2025
Units affected
2 products
Agency
FDA
Country
United States

The hazard

Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, Part Number 5867474-03; Medical Display, cathode ray tube

GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube

Brand / company
Wipro GE Healthcare Private Ltd.
Codes / lots
GTIN 00195278343383, Serial Numbers: STV25225456SA, STV25225455SA, STV25266464SA, STV25235817SA, STV25235816SA, STV25286682SA, STV25225594SA, STV25235792SA, STV25235774SA, STV25225550SA, STV25235626SA, STV25235625SA, STV25184592SA, STV25235628SA, STV25235631SA, STV25184595SA, STV25235629SA, STV25235632SA, STV25235630SA, STV25276515SA, STV25225510SA, STV25276516SA, STV25225505SA, STV25225508SA, STV25225509SA, STV25225507SA, STV25225506SA, STV25113369SA, STV25204986SA, STV25205160SA, STV25205151SA, STV25205161SA, STV25062736SA, STV25256133SA, STV25225512SA, STV25296702SA, STV25296701SA, STV25296748SA, STV25256214SA, STV25225511SA, STV25256215SA, STV25256199SA, STV25256216SA, STV25256202SA, STV25225497SA, STV25225514SA, STV25225498SA, STV25256131SA, STV25256129SA, STV25256128SA, STV25296747SA, STV25256132SA, STV25256362SA, STV25225494SA, STV25225513SA, STV25296697SA, STV25256153SA, STV25296745SA, STV25256130SA, STV25256134SA, STV25256198SA, STV25296749SA, STV25296698SA, STV25256112SA, STV25256111SA, STV25225495SA, STV25256127SA, STV25256197SA, STV25296700SA, STV25296699SA, STV23440159SA, STV23440160SA, STV23440161SA, STV23440162SA, STV23440163SA, STV23440164SA, STV23440165SA, STV23440
Distribution
Worldwide distribution.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 29 Oct 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.