The hazard
The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots identified by the firm on the IMMULITE 2000/IMMULITE 2000 XPi systems. The bias is observed in both serum and plasma samples.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Catalog Number: (1) L2KPP2, (2) L2KPP6;
- Brand / company
- Siemens Healthcare Diagnostics, Inc.
- Codes / lots
- 1. Siemens Material Number (SMN): 10381441; Catalog Number: L2KPP2; UDI-DI: 00630414961910; Lot Number: 385, ***Expanded 2/18/2026*** 387, 388, 389; 2. Siemens Material Number (SMN): 10381442; Catalog Number: L2KPP6; UDI-DI: 00630414961927; Lot Number: 385, ***Expanded 2/18/2026*** 387, 388, 389;
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, CA, DC, FL, ID, IN, LA, MD, NC, NJ, NY, PA, SC and the countries of Argentina, Australia, Bolivia, Bosnia Herzeg, Brazil, Bulgaria, Canada, Chile, Colombia, Cyprus, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, Hungary, India, Iran, Israel, Italy, Kosovo, Kuwait, Latvia, Lebanon, Macedonia, Mexico, Nepal, Netherlands, P.R. China, Pakistan, Palestinian Ter, Paraguay, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Serbia, Slovakia, Slovenia, South Africa, Spain, Taiwan, Tunisia, Turkey, Turkmenistan, U.A.E., Ukraine, Uruguay, Uzbekistan.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 5 Nov 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.