PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Zenition 50

Read the official FDA notice

Published
29 Oct 2025
Units affected
29 units
Agency
FDA
Country
United States

The hazard

Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system

Brand / company
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Codes / lots
Product Number (REF): 718096; UDI-DI: 00884838091535; All serial numbers;
Distribution
Distribution US states of MO and TX; Belgium, Canada, Fiji, Germany, India, Malaysia, Mexico, Morocco, South Africa, South Korea, Thailand, United Kingdom, and Uzbekistan.
Recall status
Ongoing