The hazard
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;
- Brand / company
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Codes / lots
- Model Number: 722234;
- Distribution
- Domestic: AL, AR, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, MD, MN, MO, NM, NY, OH, OK, PA, RI, TX, UT, VA, WA, WI, WV, WY; International: Australia, Austria, Belgium, Bulgaria, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Iraq, Italy, Japan, Netherlands, Norway, Philippines, Poland, Romania, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, United Kingdom, Vietnam;
- Recall status
- Ongoing