The hazard
The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
GM85 Digital Mobile X-ray imaging System; Model Number: GM85;
- Brand / company
- NeuroLogica Corporation
- Codes / lots
- Model Number: GM85; UDI-DI: 08806088582412; Serial Numbers: 514GM3IK400001L, 5143M3IK300001X, 5143M3IJ100010M, 5143M3IHC00014H, 5143M3IK300008R, 5143M3JJ400007E, 5143M3HJ900002V, 5143M3IK400004F, 513YM3IK500003Z, 5143M3HK600005V, 5143M3JJ200010J, 5143M3HK600008M, 5143M3HK200004H, 5143M3JK500003W, 5143M3IJ200002F, 5143M3IHC00007A, 5143M3JJ300001K, 5143M3IJ600014H, 5143M3JK500006X, 5143M3JJ200017Y, 5143M3JK400014E, 5143M3IJ600013B, 5143M3GK500004A, 5143M3IHC00011V, 5143M3JK500001M, 5143M3JJ300012N, 5143M3GK500005M, 5143M3IK400007W, 5143M3JK400008N, 5143M3JJ300011Z, 5143M3JK100003Z, 5143M3JK500007L, 5143M3IK500001F, 5143M3IHC00002L, 5143M3IK300006T, 5143M3IHC00010R, 5143M3IK600004K, 5143M3IJ300003E, 5143M3IHC00003Y, 5143M3JJ400006K, 5143M3HJ900001L, 5143M3IJ600003L, 5143M3IJC00001Y, 5143M3IJ600001H, 5143M3HJ500003A, 5149M3IK400003E, 5143M3IJ800008L, 5143M3JJ700002R, 5143M3HK300002V, 5143M3JJ200008B, 513YM3IK500002R, 5143M3JK100005D, 5143M3IJ600004F, 5143M3IHC00005J, 5143M3JJC00004Y, 5143M3HK600003R, 5143M3JK500004D, 5143M3HK600004A, 5143M3HK300012L, 5143M3HK300008M, 5143M3JJ200015M, 5143M3HK300007N, 5143M3JK100007F, 5143M3IHC00001H, 5143M3IJ400006V, 5143M3IJ100002P, 514GM3HK100003J, 5
- Distribution
- Domestic: US Nationwide Distribution. International: Australia, Belgium, Canada, Chile, China Mainland, Czech Republic, Denmark, Ecuador, Finland, France, Germany, GHANA, Hungary, India, Italy, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Spain, Sweden, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab Emir., Vietnam,
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 19 Nov 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.