NeuroLogica Corporation recalls GM85 Digital Mobile X-ray imaging System

Read the official FDA notice

Published
19 Nov 2025
Units affected
236 units
Agency
FDA
Country
United States

The hazard

The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

GM85 Digital Mobile X-ray imaging System; Model Number: GM85;

Brand / company
NeuroLogica Corporation
Codes / lots
Model Number: GM85; UDI-DI: 08806088582412; Serial Numbers: 514GM3IK400001L, 5143M3IK300001X, 5143M3IJ100010M, 5143M3IHC00014H, 5143M3IK300008R, 5143M3JJ400007E, 5143M3HJ900002V, 5143M3IK400004F, 513YM3IK500003Z, 5143M3HK600005V, 5143M3JJ200010J, 5143M3HK600008M, 5143M3HK200004H, 5143M3JK500003W, 5143M3IJ200002F, 5143M3IHC00007A, 5143M3JJ300001K, 5143M3IJ600014H, 5143M3JK500006X, 5143M3JJ200017Y, 5143M3JK400014E, 5143M3IJ600013B, 5143M3GK500004A, 5143M3IHC00011V, 5143M3JK500001M, 5143M3JJ300012N, 5143M3GK500005M, 5143M3IK400007W, 5143M3JK400008N, 5143M3JJ300011Z, 5143M3JK100003Z, 5143M3JK500007L, 5143M3IK500001F, 5143M3IHC00002L, 5143M3IK300006T, 5143M3IHC00010R, 5143M3IK600004K, 5143M3IJ300003E, 5143M3IHC00003Y, 5143M3JJ400006K, 5143M3HJ900001L, 5143M3IJ600003L, 5143M3IJC00001Y, 5143M3IJ600001H, 5143M3HJ500003A, 5149M3IK400003E, 5143M3IJ800008L, 5143M3JJ700002R, 5143M3HK300002V, 5143M3JJ200008B, 513YM3IK500002R, 5143M3JK100005D, 5143M3IJ600004F, 5143M3IHC00005J, 5143M3JJC00004Y, 5143M3HK600003R, 5143M3JK500004D, 5143M3HK600004A, 5143M3HK300012L, 5143M3HK300008M, 5143M3JJ200015M, 5143M3HK300007N, 5143M3JK100007F, 5143M3IHC00001H, 5143M3IJ400006V, 5143M3IJ100002P, 514GM3HK100003J, 5
Distribution
Domestic: US Nationwide Distribution. International: Australia, Belgium, Canada, Chile, China Mainland, Czech Republic, Denmark, Ecuador, Finland, France, Germany, GHANA, Hungary, India, Italy, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Spain, Sweden, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab Emir., Vietnam,
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 19 Nov 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.