The hazard
Acute circulatory support system pump may not fit into the motor and reports received of the pump not being securely locked into the motor using the screw-in locking feature. Failure to properly lock the pump into motor my not cause pump not inserted alarm, and could lead to particulate, increased heat, anxiety, inconvenience, delay in hemodynamic/ECMO support, stroke, embolism, death.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre-connected Pack, REF: CMAEK00; CentriMag VAD KIT, REF: 201-51006, 201-51007, 201-90001 201-90001T; Blood Pump with CentriMag Acute Circulatory Support System for ECMO, REF: CMAEK01. CentriMag Blood Pump is a component to be used with CentriMag Acute Circulatory Support System
- Brand / company
- Thoratec LLC
- Codes / lots
- UDI-DI: 07640135140627, 07640135140603, 05415067036414 IFU: Part(Name) Revision: For US: 1. ARTEN600254920 (IFU CentriMag Blood Pump ABT) Rev C 2. ARTEN600254958 (IFU PediMag Blood Pump US ABT) Rev C 3. ARTEN100174794 (CentriMag Adult Pre-connected Pack IFU, US) Rev B 4. ARTEN600230157 (IFU CentriMag Blood Pump ECMO US) Rev D FOR OUS: ARTEN600254935 (IFU CentriMag Blood Pump CAN ABT) Rev C ARTEN600254936 (IFU PediVas Blood Pump Canada ABT) Rev C ARTEN600332205 (IFU CMag VAS-ECMO Pump OUS) Rev A ARTEN600332207 (IFU PVAS VAS-ECMO Pump OUS) Rev A REF 201-20003, 104318, 201-51006, 201-51007, 201-90001 201-90001T don't have associated un-expired batches/serial numbers. REF/Batch(Serial): 102953/10007334, 10007335, 10007336, 10012012, 10012013, 10012014, 10012015, 10012016, 10036179, 10036180, 10036181, 10037990, 10037992, 10037993, 10037995, 10037996, 10046028, 10046029, 10046030, 10046031, 10046032, 10046033, 10058754, 10058755, 10058756, 10061673, 10061675, 10061676, 10061677, 10065946, 10065947, 10065954, 10065955, 10065956, 10071358, 10071359, 10071361, 10073780, 10073781, 10073783, 10082053, 10082055, 10103208, 10103209, 10106345, 10106346, 10106347, 10106348, 10120074, 10120075, 1
- Distribution
- Worldwide - US Nationwide distribution including in the states of MN, PA, OK, FL, CA, OR, IL, NY, DE, GA, NJ, NC, WI, OH, AZ, AR, TX, TN, MO, MA, MT, KY, NE, WV, VA, ME, ND, AL, CO, DC, MS, IN, MI, CT, SC, WA, NM, MD, HI, ID, IA, LA, UT, RI, NV and the countries of Algeria, Australia, Austria, Bahrain, CANADA, Cayman Islands, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Hong Kong, Italy, Jordan, Kuwait, Lithuania, Macedonia, Malaysia, Mexico, Netherlands, Oman, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 21 Jan 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.