The hazard
There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
MAGNETOM Cima.X Upgrade. Model Number: 11689304.
- Brand / company
- Siemens Medical Solutions USA, Inc
- Codes / lots
- Model Number: 11689304. UDI: (01)04056869975245(21)237304. Serial Number: 237304.
- Distribution
- US distribution to California. International distribution to Australia, Germany, Great Britian.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 19 Nov 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.