The hazard
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
MAMMOMAT Inspiration;
MAMMOMAT Fusion;
MAMMOMAT Revelation;
- Brand / company
- Siemens Medical Solutions USA, Inc
- Codes / lots
- Model Number: 10140000; UDI-DI: 04056869009063; System Serial Numbers: 6734, 6694, 10121, 10309, 10336, 10342; | Model Number: 10140000; UDI-DI: 04056869009063; System Serial Numbers: 10559; | Model Number: 11343300; UDI-DI: 04056869124575; System Serial Numbers: 1296, 10550, 10572, 10606, 10608, 11511, 20140;
- Distribution
- Worldwide - US Nationwide distribution in the states of AR, CA, NY and the countries of Australia, Canada, India.
- Recall status
- Ongoing