CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion Systems…

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion Systems with software versions R2.1.10 and R2.2.10

Read the official FDA notice

Published
21 Jan 2026
Units affected
10,466 units
Agency
FDA
Country
United States

The hazard

Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Philips Azurion Systems with software versions R2.1.10 and R2.2.10. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Numbers: 722063, 722221. 2. Azurion 3 M15. Model Numbers: 722064, 722222, 722280. 3. Azurion 5 M12. Model Number: 722227. 4. Azurion 5 M20. Model Numbers: 722228, 722281. 5. Azurion 7 B12. Model Numbers: 722067, 722225. 6. Azurion 7 B20. Model Numbers: 722068, 722226. 7. Azurion 7 M12. Model Numbers: 722078, 722223. 8. Azurion 7 M20. Model Numbers: 722079, 722224, 722282.

Brand / company
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Codes / lots
Software Versions: R2.1.10 and R2.2.10. 1. Model Number (UDI): 722063 ((01)00884838085275(21)122, (01)00884838085275(21)64, (01)00884838085275(21)56, (01)00884838085275(21)131, (01)00884838085275(21)84, (01)00884838085275(21)132, (01)00884838085275(21)96, (01)00884838085275(21)98, (01)00884838085275(21)108, (01)00884838085275(21)153, (01)00884838085275(21)69, (01)00884838085275(21)154, (01)00884838085275(21)97, (01)00884838085275(21)137, (01)00884838085275(21)116, (01)00884838085275(21)127, (01)00884838085275(21)151, (01)00884838085275(21)148, (01)00884838085275(21)150, (01)00884838085275(21)146, (01)00884838085275(21)149, (01)00884838085275(21)123, (01)00884838085275(21)130, (01)00884838085275(21)143, (01)00884838085275(21)139, (01)00884838085275(21)144, (01)00884838085275(21)145, (01)00884838085275(21)141, (01)00884838085275(21)140, (01)00884838085275(21)135, (01)00884838085275(21)133, (01)00884838085275(21)107, (01)00884838085275(21)120, (01)00884838085275(21)129, (01)00884838085275(21)105, (01)00884838085275(21)110, (01)00884838085275(21)128, (01)00884838085275(21)119, (01)00884838085275(21)113, (01)00884838085275(21)43, (01)00884838085275(21)117, (01)00884838085275(21)115, (01
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine/United Arab Emirate, United Kingdom, Uruguay, Uzbekistan, Viet Nam, Vietnam, Yemen.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 21 Jan 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.