AGFA Healthcare Corp. recalls Brand name

Read the official FDA notice

Published
31 Dec 2025
Units affected
35
Agency
FDA
Country
United States

The hazard

It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800

Brand / company
AGFA Healthcare Corp.
Codes / lots
N/A
Distribution
U.S. Nationwide distribution.
Recall status
Ongoing