CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Maquet Cardiopulmonary Gmbh recalls Bubble Sensor for 3/8" x 3/32"…

Maquet Cardiopulmonary Gmbh recalls Bubble Sensor for 3/8" x 3/32" tubing

Read the official FDA notice

Published
11 Feb 2026
Units affected
3050 units
Agency
FDA
Country
United States

The hazard

Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device. These errors can occur temporarily when the connecting cable is moved or permanently if the connection is fully compromised.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;

Brand / company
Maquet Cardiopulmonary Gmbh
Codes / lots
Model Number: 701055720; UDI-DI: 04037691816432; Serial Numbers: All units impacted;
Distribution
Worldwide Distribution: US (Nationwide) and OUS (International) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, Estonia, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Korea, Republic of (South Korea), Latvia, Lithuania, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, Uzbekistan, and Vietnam
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 11 Feb 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.