CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Medtronic MiniMed, Inc. recalls MiniMed 720G Insulin Pump (MMT-1809…

Medtronic MiniMed, Inc. recalls MiniMed 720G Insulin Pump (MMT-1809 and 13 more

Read the official FDA notice

Published
8 Apr 2026
Units affected
14 products
Agency
FDA
Country
United States

The hazard

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

MiniMed 720G Insulin Pump (MMT-1809, MMT-1810, MMT-1859, MMT-1860, MMT-1867)

Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)

Paradigm REAL-Time Veo Insulin Pump (MMT-554, MMT-754)

MiniMed 620G Insulin Pump (MMT-1750)

Paradigm Insulin Pump (MMT-712, MMT-715)

MiniMed 740G Insulin Pump (MMT-1811, MMT-1812, MMT-1861, MMT-1862)

Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)

MiniMed 530G Insulin Pump (MMT-551, MMT-751)

MiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)

MiniMed 640G Insulin Pump (MMT-1711, MMT-1712, MMT-1751, MMT-1752)

MiniMed 630G Insulin Pump (MMT-1714, MMT-1715, MMT-1754, MMT-1755)

MiniMed 770G Insulin Pump (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)

Brand / company
Medtronic MiniMed, Inc.
Codes / lots
Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 720G Insulin Pump: 763000854768, 763000505196, 763000505219, 763000505318, 763000505257, 763000505295, 763000505233, 763000505370, 763000505325, 763000505332, 763000505394, 763000505356, 763000505387, 763000505349, 763000505400, 763000505226, 763000505271, 763000638016, 763000531171, 763000854751: MINIMED 720G SYSTEM USER GUIDE(M000187C022 1, M000186C002 1, M000186C014 1, M000187C016 1, M000186C021 1, M000186C018 1, M000186C017 1, M000187C010 1, M000187C011 1, M000186C031 1, M000187C013 1, M000187C008 1, M000187C015 1, M000187C005 1, M000187C033 1, M000186C028 1, M000187C023 1, M000187C020 1, M000187C018 1, M000186C018 13, M000186C004 1, M000186C030 1, M000186C009 1, M000187C024 1, M000186C033 1, M000186C003 1, M000187C012 1). | Pump: UDI-DI(GTIN): User Guide Name(Part-Version): Paradigm Real-Time Insulin Pump: 00613994409423, 00613994622600, 613994622747, 613994407733, 00613994407740, 00613994136374, 00613994271600, 00613994106476, 00613994373120, 00613994409478, 00613994312426, 00613994151780, 00885074419831, 00613994622747, 00613994951502, 00613994622648, 00643169026100, 00613994622624, 00613994622556, 00613994402011, 00
Distribution
Worldwide - US Nationwide distribution including in the states of OR, CA, AZ, NY,TX,FL, OH, MO, TN,NJ, NC, SC, IL, VA, WA, DC, GA, PA, MS,IA, MI, NH, OK, SD, MT, WI, WV, MA, LA, MN, RI, AR, ME, AL, IN, UT, KY, CO, ND, MD, CT, NM, KS, ID, NV, DE, NE, VT, AK, HI, WY, PR, VI and the countries of Taiwan, New Zealand, Australia, Bangladesh, Brunei Darussalam, Cambodia, Indonesia, Malaysia, Maldives, Philippines, Thailand, Vietnam, Singapore, Pakistan, Hong Kong, China, Argentina, Aruba, Barbados, Bermuda, Bolivia, Plurinational State Of, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, Puerto Rico, Trinidad And Tobago, Uruguay, Venezuela, Canada, South Korea, India, Albania, Algeria, Andorra, Armenia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Bulgaria, Canary Islands, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guyana, French Polynes., Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Macedonia, Malawi, Malta, Martinique, Mauritius, Mayotte, Moldavia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Qatar, Reunion, Romania, Russia, San Marino, Saudi Arabia, Serbia, Sierra Leone, Slovakia, Slovenia, South Africa, Spain, St.Pier,Miquel., Sudan, Sweden, Switzerland, Syrian Arab Republic, Tanzania, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Yemen, Zambia, Zimbabwe.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 8 Apr 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.